Wednesday, November 26, 2008

Praning5254 Shielding and X-ray Room Design

Purpose of Shielding

To protect :

  • the x-ray department staff
  • the patients (when not being x-rayed)
  • visitors and the public
  • persons working adjacent to or near the x-ray facility

Radiation Shielding - Design Concepts

Data required include consideration of :

  • Type of x-ray equipment
  • Usage (workload)
  • Positioning
  • Whether multiple tubes/receptors are being used
  • Primary beam access (vs. scatter only)
  • Operator location
  • Surrounding areas

Shielding Design (I)

Equipment

What equipment is to be used ?

  • General radiography
  • Fluoroscopy (with or without radiography)
  • Dental (oral or OPG)
  • Mammography
  • CT

Shielding Design (II)

The type of equipment is very important for the following reasons :

  • where the x-ray beam will be directed
  • the number and type of procedures performed
  • the location of the radiographer (operator)
  • the energy (kVp) of the x-rays

Shielding Design (III)

Usage

  • Different x-ray equipment have very different usage.
  • For example, a dental unit uses low mAs and low (~70) kVp, and takes relatively few x-rays each week
  • A CT scanner uses high (~130) kVp, high mAs, and takes very many scans each week.

Shielding Design (IV)

  • The total mAs used each week is an indication of the total x-ray dose administered
  • The kVp used is also related to dose, but also indicates the penetrating ability of the x-rays
  • High kVp and mAs means that more shielding is required.

Shielding Design (V)

Positioning (...Continued on the next post)

0 comments:

  © Blogger template 'External' by Ourblogtemplates.com 2008

Back to TOP